  • Validation Data

1. Yield and Purity of Spin Columns

 Sample       Column  Conc.,ng/μL  m,μg  A260/A280  A260/A230
 Customer A Buffer  NP30-A  93  9.3  1.8  2
 95  9.5  1.9  2.1
 90  9  1.8  2
 Customer A Buffer  Qiagen Miniprep 
 column
 
 83  8.3  1.8  2
 88  8.8  1.8  2.1
 90  9  1.8  2
 QIAprep Spin Miniprep Kit
 Qiagen, Cat. No 27104
 NP30-A  100  10  1.8  2
 106  10.6  1.9  2
 104  10.4  1.9  2.1
 QIAprep Spin Miniprep Kit
 Qiagen, Cat. No 27104
 Qiagen Miniprep 
 column
 
 102  10.2  1.9  2
 101  10.1  1.8  2
 102  10.2  1.9  2
Results: The performance of NP30-A is equivalent to Qiagen spin columns with both customer and Qiagen buffer.

 

2. Stability of Spin Columns

 Sample       Column  Conc.,ng/μL  m,μg  A260/A280  A260/A230
 NP30-A: 1 year of storage  Customer A Kit  141  14.1  1.9  2
 147  14.7  1.8  2.1
 151  15.1  1.9  2.2
 NP30-A: New lot.  Customer A Kit  142  14.2  1.8  2.2
 159  15.9  1.8  1.8
 149  14.9  1.8  2.3
 NP30-A: 1 year of storage  QIAprep Spin 
 Miniprep Kit
 Qiagen, Cat.
 No 27104
 168  16.8  1.8  2.0 
 178  17.8  1.9  2.1
 160  16.0   1.8  2.0 
 NP30-A: New lot.  QIAprep Spin 
 Miniprep Kit
 Qiagen, Cat.
 No 27104
 162  16.2  1.8  2.1
 169  16.9  1.8  2.1
 169  16.9  1.9  2.0 
Results: NP30-A show good performance after one year storage.

 

3. Purity Testing of Silica Membrane

 Membrane  Extractant  m/z  Ion  Compound  Source
 Biocomma Silica
 Membrane
 Pure Water  304  M+Na+  oleamide  Slip agent in polyethylene films
 97  HSO4-    
 H2PO4-    
 Methanol  360  M+Na+  erucamide  Slip agent in polyethylene films
 Filters from
 Well-known brand
 Pure Water  304  M+Na+  oleamide  Slip agent in polyethylene films
 97  HSO4-    
 H2PO4-    
 Methanol  360  M+Na+  erucamide  Slip agent in polyethylene films
 425  M+Na+  Unidentified contaminant  
 701  M+Na+    from water and solvent filters

 

Remarks:
1. The table only records ions that match the “ESI Common Background Ions” reference table.
2. The results have deducted the ions present in the blank, but if a particular m/z ion significantly exceeds the blank, it is recorded.
Conclusion:
Proceed ultrasonic extraction in both water and methanol, mass spectrometry results showed the presence of precipitates in both Biocomma membrane and well-known brand filter. Compound identification was carried out through high-resolution mass spectrometry by full scanning and matching with the “ESI Common Background Ions” reference table. The primary precipitates were HSO4-, H2PO4-, and Na+, and these presence is attributed to the material being glass fiber in the membranes. No other substances were found to precipitate from either of the two membranes, confirming their purity.