Copure® HPLC Columns Specially for Pesticide Residue
Copure® HPLC Columns Specially for Pesticide Residue utilize ultra-high purity silica with narrow particle size distribution, combined with mature and stable bonding technology and precise end-capping process, ensuring high column efficiency and a suitable pressure range. In pesticide residue analysis, it provides high sensitivity, excellent separation, and symmetrical peak shapes, thereby enhancing accuracy and reliability in detection.
Details
Instrument: UPLC-MS/MS (Thermo Fisher TSQ Endura)
Column: Copure® HPLC Columns Specially for Pesticide Residue (2.1 mm×100 mm, 3 μm)
Part Number: HNC01
Mobile Phase: A: Ammonium Formate-Formic Acid Water
(2mmol/L) B: Ammonium Formate-Formic Acid Methanol
(2mmol/L)
Flow Rate: 0.3 mL/min
Column Temperature: 40℃
Injection Volume: 2 μL
Elution Mode: Gradient Elution, see Table 1
Table 1: Gradient Elution Program
| Time /min |
A/% |
B/% |
| 0.00 |
97 |
3.0 |
| 1.00 |
97 |
3.0 |
| 1.50 |
85 |
15.0 |
| 2.50 |
50 |
50.0 |
| 18.0 |
30 |
70.0 |
| 23.0 |
2 |
98.0 |
| 25.0 |
2 |
98.0 |
| 25.1 |
97 |
3.0 |
| 27.0 |
97 |
3.0 |
4.2 Mass Spectrometry Conditions
Ion Source: HESI
Electrospray Voltage: Positive Ion Mode 3500 V, Negative Ion Mode 3500V
Sheath Gas Pressure: 35 arb
Auxiliary Gas Pressure: 3 arb
Ion Transfer Tube: 380℃
Auxiliary Temperature: 420℃

Metalaxyl

Methomyl Sulfoxide

Aldicarb Sulfoxide

Methomyl

Thiamethoxam

Imidacloprid

Clothianidin

Acetamiprid

Thiophanate-methy

Chlorfenapyr

Carbendazim

3-Hydroxycarbofuran

Terbufos

Carbofuran

Azoxystrobin

Dimethomorph

Fluoromethane

Prochloraz - Desimidazolecarbamoyl

Prochloraz - Dechlorinated imidazole

Prochloraz

Flusilazole

Fipronil

Fipronil sulfide

Phoxim

Pyraclostrobin

Oxathiapiprolin

Emamectin Benzoate

Pyridaben

Abamectin

Fipronil sulfone

Cyantraniliprole

Etoxazole
Ordering Information
| Cat.# |
Specification |
Qty. |
| HNC01 |
2.1mm×100mm,3μm |
1 unit/box |
| HNC02 |
2.1mm×100mm,1.7μm |
1 unit/box |
Write a comment